Related Experiment Videos
Locking mechanism failure in semiconstrained total elbow arthroplasty
Mark P Figgie1, Edwin P Su, Barbara Kahn
1Comprehensive Arthritis Service, Hospital for Special Surgery, 535 E. 70th Street, New York, NY 10021, USA. figgiem@hss.edu
Journal of Shoulder and Elbow Surgery
|January 18, 2006
Summary
Late failure of semiconstrained total elbow implants occurred in 5.3% of cases, with a mean failure time of 5.5 years. Revision surgery was successful in most patients, restoring function and motion.
Area of Science:
- Orthopedic Surgery
- Biomedical Engineering
- Implant Design
Background:
- Total elbow replacement is a common procedure for end-stage elbow arthritis.
- Semiconstrained implants offer stability but can be prone to mechanical failure.
- Locking mechanism failure can lead to significant morbidity and revision surgery.
Purpose of the Study:
- To evaluate the incidence and outcomes of late locking mechanism failure in semiconstrained total elbow replacements.
- To assess the efficacy of a novel revision technique for polyethylene bushing and locking mechanism failure.
Main Methods:
- Retrospective review of 170 total elbow replacements performed between 1988 and 2000 using semiconstrained implants.
- Analysis of radiographic evidence for implant failure, specifically locking mechanism migration.
- Surgical revision of failed locking mechanisms using a custom-made locking clip via a limited approach.
Main Results:
- A 5.3% (9 of 170) incidence of late locking mechanism failure was observed, with a mean time to failure of 5.5 years.
- Eight elbows required revision surgery; 7 of these had successful revisions with no recurrence at a mean follow-up of 1.4 years.
- Revised implants in 7 patients demonstrated good or excellent outcomes using the Hospital for Special Surgery elbow scoring system, with a mean arc of motion of 130 degrees.
Conclusions:
- Late locking mechanism failure is a potential complication of semiconstrained total elbow arthroplasty.
- A limited surgical approach with a custom locking clip can effectively revise polyethylene bushing and locking mechanism failures.
- Successful revision surgery can restore good to excellent functional outcomes and range of motion in most patients.