Quantifying the effect of IV contrast media on integrated PET/CT: clinical evaluation
Osama Mawlawi1, Jeremy J Erasmus, Reginald F Munden
1Department of Imaging Physics, The University of Texas M. D. Anderson Cancer Center, 1515 Holcombe Blvd., Box 56, Houston, TX 77030, USA. Omawlawi@mdanderson.org
Objectives:
The use of IV contrast media in PET/CT can result in an overestimation of PET attenuation factors that potentially can affect interpretation. The objective of this study was to quantify the effect of IV contrast media in PET/CT and assess its impact on patients with intrathoracic malignancies.
Materials And Methods:
Nine patients had CTs performed with and without IV contrast media followed by (18)F-FDG PET. PET images were reconstructed using contrast-enhanced and unenhanced CT. To quantify the effect of contrast media on standardized uptake values (SUV), similar regions of interest (ROIs) were drawn on the subclavian vein, heart, liver, spleen, and site of malignancy on both CT and corresponding reconstructed PET images, and the mean and maximum values were compared. In addition, two physicians blinded to the imaging parameters that were used evaluated the reconstructed PET images to assess whether IV contrast media had an effect on clinical interpretation.
Results:
For all patient studies, the subclavian vein region on the ipsilateral side of contrast media administration had the highest increase in CT numbers with a corresponding average SUV(max) increase of 27.1%. Similarly, ROIs of the heart and at the site of malignancy showed an increase in the maximum attenuation value with a corresponding average SUV(max) increase of 16.7% and 8.4%, respectively. Other locations had relatively small attenuation value differences with a correspondingly negligible SUV variation.
Conclusion:
Although there is a significant increase in SUV in regions of high-contrast concentration when contrast-enhanced CT is used for attenuation correction, this increase is clinically insignificant. Accordingly, in PET/CT, IV contrast-enhanced CT can be used in combination with the PET to evaluate patients with cancer.
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies VII: Vascular Imaging