Comparison of resolution, contrast, and color differentiation among fiberoptic and digital flexible cystoscopes
James F Borin1, Corollos S Abdelshehid, Ralph V Clayman
1Department of Urology, University of California, Irvine, Orange, California 92868, USA.
Background And Purpose:
Advances in electro-optics continue to improve the urologist's ability to perform minimally invasive procedures. While the development of flexible fiberoptic cystoscopes more than 20 years ago greatly impacted the practice of urology, distal-sensor digital technology may represent the next step in the evolution of endoscopy. We compared a new distal-sensor digital flexible cystoscope with two standard fiberoptic flexible cystoscopes.
Materials And Methods:
We evaluated the resolution, contrast, and color discrimination of a new ACMIICN distal-sensor digital cystoscope with >165,000 effective pixels in its clear aperture (viewing area), a new ACMI-ACN II fiberoptic cystoscope with <15,000 pixels in its clear aperture, and a Storz 1127 office fiberoptic cystoscope. Five subjects compared each cystoscope across 13 test parameters.
Results:
There was no difference in the performance of the two fiberoptic cystoscopes. The ICN cystoscope was statistically superior to one or both fiberoptic cystoscopes across 12 of the 13 tests, including color differentiation between shades of dark red (P < 0.05), contrast discrimination along a 15-step grayscale gradient (P < 0.001 compared with the 1127 fiberoptic cystoscope only), resolution at 10 mm (7.52 line pairs/mm [lp/mm] (ICN) nu 3.58 lp/mm for both fiberoptic cystoscopes; (P < 0.001), and clear resolution of a 1-mm target at a distance of 6.1 cm (ICN) nu 3.3 cm (1127) and 3.8 cm (ACN II) (P < 0.001).
Conclusions:
The ICM distal-sensor all-digital cystoscope was clearly superior to two representative fiberoptic cystoscopes in vitro in terms of resolution, contrast discrimination, and red color differentiation. In-vivo performance remains to be assessed.
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Confocal Fluorescence Microscopy
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):