Related Experiment Video
Updated: Aug 13, 2026

Isolation of High-density Lipoproteins for Non-coding Small RNA Quantification
Published on: November 28, 2016
An assay for separating and quantifying four bilirubin fractions in untreated human serum using isocratic
Susumu Osawa1, Shin Sugo, Toshihiko Yoshida
1Division of Biological Sciences and Technology, Department of Health Sciences, School of Medicine, Kyushu University, 3-1-1, Maidashi, Higasshi-Ku, Fukuoka City, 812-8582, Japan. osawas@shs.kyushu-u.ac.jp
Background:
The quantification of serum bilirubin fractions has been widely performed with both the diazo-method and an enzymatic method; however, the accuracy of these methods has not been evaluated because quantitative fractional high-performance liquid chromatography (HPLC) reference methods have yet to be established.
Methods:
Samples were analyzed using HPLC and Shodex Asahipak GS-320HQ columns. Human serum was subjected to HPLC using direct injection, then eluted with acetonitrile: 0.3 mol/l phosphate buffer (pH 6.5) containing 1% Brij 35 and 0.08% sodium ascorbate (30:70, v/v).
Results:
Serum bilirubin was separated into 4 fractions; retention times of 9.24, 19.92, 24.07, 35.75 min were identified as delta bilirubin, bilirubin diglucuronide, bilirubin monoglucuronide, and unconjugated bilirubin, respectively. Mean recovery was 93.0%-99.2%. Total precision of peak retention time, height and area exhibited <4.26% variation. Detection range was 3.1 to 348 mg/l. Hemoglobin (6 g/l) and immunoglobins produced a small positive interference. beta-carotene (20 mg/l), vitamin-B2 (370 microg/l) and B(12) (9.5 microg/l) did not interfere with this analysis. Results (n=30) with this method were closely correlated to those by Adachi's HPLC method as r=0.9941 to 0.9960, slope=0.88 to 1.27, intercept=-3.2 to +4.9, for each fraction.
Conclusions:
Since this method was a precise quantitative HPLC method for serum bilirubin fractionation, it might be used to evaluate the accuracy and the characteristics of various routine methods for bilirubin measurement.
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Affinity Chromatography