Related Experiment Video
Updated: Aug 13, 2026

Surgical Porcine Model of Chronic Myocardial Ischemia Treated by Exosome-laden Collagen Patch and Off-pump Coronary Artery Bypass Graft
Published on: September 15, 2023
Portable coronary active perfusion system for off-pump coronary artery bypass grafting
Yoshinao Koshida1, Go Watanabe, Tamotsu Yasuda
1Department of General and Cardiothoracic Surgery, Kanazawa University School of Medicine, Kanazawa, Ishikawa, Japan. yoshinaoko2000@yahoo.co.jp
Purpose:
The present study was performed to develop a new perfusion system for off-pump coronary artery bypass grafting and to examine whether even a simple coronary perfusion system can maintain adequate blood flow delivery during anastomosis.
Description:
The experiment was performed in two stages. In procedure 1, 3 pigs with left anterior descending artery occlusion were used to evaluate optimal perfusion flow rate and coronary artery internal pressure, and to evaluate the safety area of perfusion. In procedure 2, 6 pigs were used to validate the new portable coronary perfusion system.
Evaluation:
The optimal blood flow in the portable coronary active perfusion system was less than approximately 40 mL/min. The small, easy to use pump system (ie, the portable coronary active perfusion system) may prevent hemodynamic deterioration and ventricular arrhythmia during coronary occlusion, resulting in better maintenance of left ventricular function.
Conclusions:
Even a simple pump system can achieve effective perfusion for safe anastomosis. Further studies are required to allow the clinical use of this system.