The OxyArm: a supplemental oxygen delivery device.

James W Futrell1, Jack L Moore

  • 1Department of Anesthesiology, Cedars-Sinai Medical Center, Los Angeles, California, USA.

Anesthesia and Analgesia
|January 24, 2006
PubMed
Summary

The new OxyArm device provides supplemental oxygen to patients post-anesthesia without facial contact, maintaining safe oxygen saturation levels. This open oxygen system offers comfort and ease of use, improving patient care during the early recovery period.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same journal

Perioperative Interventions in Autism Spectrum Disorder: A Systematic Review of Randomized Trials.

Anesthesia and analgesia·2026
Same journal

Early Adoption of Suzetrigine After FDA Approval: Inpatient and Outpatient Practice Patterns.

Anesthesia and analgesia·2026
Same journal

Clinical Effectiveness of Acute Normovolemic Hemodilution in Cardiac Surgery: A Systematic Review and Meta-analysis.

Anesthesia and analgesia·2026
Same journal

The Next Frontier in Anesthesia Safety-From Standards to Implementation: APSF Ellison C. Pierce Lecture 2025 and WCA Harold Griffith Lecture 2026.

Anesthesia and analgesia·2026
Same journal

Multimodal Machine Learning Model Predicting Postoperative Delirium Based on Heart Rate Variability: A Prospective Observational Study.

Anesthesia and analgesia·2026
Same journal

Feasibility of Application of Artificial Intelligence in Neuroanesthesia and Neurocritical Care in Low- and Middle-Income Countries.

Anesthesia and analgesia·2026
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Area of Science:

  • Anesthesiology
  • Respiratory Therapy
  • Medical Device Innovation

Background:

  • Traditional supplemental oxygen delivery methods like facemasks and nasal cannulae can cause patient complications post-anesthesia, including aspiration and facial trauma.
  • Existing devices require direct facial contact, potentially interfering with patient comfort and nursing care.

Purpose of the Study:

  • To evaluate the efficacy and safety of the OxyArm, a novel open oxygen system, for supplemental oxygen delivery in the immediate postoperative period.
  • To compare oxygen saturation levels using the OxyArm with those during endotracheal intubation.

Main Methods:

  • A clinical study involving 60 patients receiving supplemental oxygen via the OxyArm for 8 minutes after general anesthesia and tracheal extubation.
  • Continuous monitoring of oxygen saturation at three time points: pre-extubation (endotracheal tube), and 4 and 8 minutes post-extubation using the OxyArm at different flow rates (4 L/min and 2 L/min).

Related Experiment Videos

Main Results:

  • No significant differences in oxygen saturation were observed among the three measurement periods.
  • No patient experienced an oxygen desaturation event below 88%.
  • Patients and clinicians reported high comfort and ease of use, noting the device facilitates facial care without interrupting oxygen therapy.

Conclusions:

  • The OxyArm effectively delivers adequate oxygen levels for most patients in the early postoperative phase.
  • The open oxygen system design eliminates complications associated with traditional devices and enhances patient comfort and care.
  • The OxyArm represents a promising alternative for postoperative oxygenation, improving patient and clinician experience.