Determination of trace amounts of cobalt(II) by flow injection-solid phase spectrometry (FI-SPS) with 5-Br-PADAB
Shiro Matsuoka1, Nobutoshi Shiota, Kazuhisa Yoshimura
1Department of Environmental Science, Faculty of Science, Niigata University, Ikarashi, Niigata 950-2181, Japan. matsuoka@env.sc.niigata-u.ac.jp
Abstract:
Flow injection-solid phase spectrometry (FI-SPS) has been applied to the determination of cobalt(II) in water samples. The complex formed between cobalt and 5-Br-PADAB was on-line protonated and concentrated on an AG 50W-X2 cation-exchanger in a flow-through cell. The increase in absorbance caused by the accumulation of the complex in the resin was continuously measured. The interference by copper could be effectively eliminated by using EDDP as a masking agent. The detection limit was 40 ng dm(-3) with a 4.0 cm3 sample solution.
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Gas Chromatography: Types of Detectors-II
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.