Related Experiment Video
Updated: Aug 13, 2026

Determination of Cardiac Output in a Porcine Model for Ex Vivo Pulmonary Perfusion
Published on: June 28, 2024
Cardiac output measurement by pulse dye densitometry: comparison with pulmonary artery thermodilution in post-cardiac
Martijn Kroon1, A B Johan Groeneveld, Yvo M Smulders
1Department of Internal Medicine, VU University Medical Center, De Boelelaan 1117, 1081 HV, Amsterdam, The Netherlands.
Objective:
Pulse-dye densitometry (PDD) could be a suitable, low-invasive alternative to thermodilution using a pulmonary artery catheter (PAC) for monitoring cardiac output. The aim of our study was to assess the reproducibility and validity of PDD compared to PAC-thermodilution.
Methods:
In 43 post-cardiac surgery patients, the mean of triplicate readings of cardiac output was assessed using both methods. In a subgroup of 26 patients, a second set of measurements was obtained on average 2 h later.
Results:
Reproducibility of consecutive measurements was slightly better for PAC-thermodilution than for PDD (median coefficient of variation of the triplicate measurements: 3.5% versus 5.4%, P < 0.01). Both methods correlated well (r = 0.84, p < 0.001). Using Bland and Altman analyses with PAC-thermodilution as the reference method, PDD showed a bias of -0.68 +/- 0.82 L/min, mainly due to differences in higher ranges of cardiac output (>6.5 L/min). Measured changes in cardiac output were 81% concordant (i.e. <1 L/min different) between both methods.
Conclusion:
PDD correlates well with PAC-thermodilution and thus deserves consideration as a low-invasive alternative for measurement and follow-up of cardiac output.
Related Concept Videos
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...