Microfabricated refractive index gradient based detector for reversed-phase liquid chromatography with mobile phase
Adam D McBrady1, Robert E Synovec
1Center for Process Analytical Chemistry, Department of Chemistry, Box 351700, University of Washington, Seattle, WA 98195-1700, USA.
This study introduces a novel microfabricated refractive index gradient (micro-RIG) detector for liquid chromatography (LC). The micro-RIG detector successfully overcomes baseline shift issues in gradient elution, enabling sensitive and reproducible separations of proteins and polymers.
Area of Science:
- Analytical Chemistry
- Chromatography
- Detector Technology
Background:
- Traditional refractive index (RI) detectors struggle with mobile phase gradient elution in liquid chromatography (LC) due to baseline drift.
- Accurate detection during gradient elution is crucial for sensitive and reproducible chromatographic separations.
Purpose of the Study:
- To develop and demonstrate a sensitive, highly reproducible microfabricated refractive index gradient (micro-RIG) detector for gradient elution LC.
- To enable robust baseline correction procedures for gradient LC separations.
Main Methods:
- Utilized a microfabricated borosilicate glass chip with a collimated diode laser beam to probe a refractive index gradient (RIG).
- Implemented a baseline correction method involving subtraction of mobile phase gradient blanks.
- Performed reversed-phase high-performance liquid chromatography (RP-HPLC) separations of proteins and polymers under varying gradient conditions.
Main Results:
- The micro-RIG detector maintained on-scale signals throughout mobile phase gradients, allowing for effective baseline correction.
- Achieved reproducible, flat baselines after blank subtraction, leading to high signal-to-noise ratios for analytes.
- Demonstrated consistent limits of detection (LODs), with a 3-sigma deflection angle LOD of 16 microrad, corresponding to 7 ppm for cytochrome c.
Conclusions:
- The chip-based micro-RIG detector is a viable solution for sensitive and reproducible gradient elution LC.
- The detector design is compatible with both traditional and miniaturized LC systems.
- Despite minor sensitivity variations due to viscosity effects, the micro-RIG detector offers reliable performance for complex separations.
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Types of Detectors
Ion-Exchange Chromatography