First clinical experience with an automatic control system for rotary blood pumps during ergometry and right-heart
Heinrich Schima1, Michael Vollkron, Ursula Jantsch
1Department of Cardiothoracic Surgery and Ludwig Boltzmann Institute for Cardiosurgical Research, Medical University of Vienna, Vienna, Austria. heinrich.schima@meduniwien.ac.at
Background:
At present, most clinically implanted rotary blood pumps are operated at constant speed and adjusted by the physician. It is generally assumed that an adaptation of pump speed to the patient's physiologic requirements would be beneficial. The data provided in this paper, based on hemodynamic and spirometric data during exercise in which a pre-load-sensitive control was used, lend quantitative support to this assumption.
Methods:
An automatic speed control was developed and implemented with Matlab on a dSpace controller board. The system uses pump speed, pump power, and pump flow as its only input signals. It was connected to the clinical hardware of the DeBakey VAD System. The control is pre-load-sensitive and uses an expert system to detect excessive unloading and eventual suction. This system was used to quantify the cardiovascular reaction of patients to both automatically controlled and constant pump speed. A sub-group of 5 patients underwent bicycle ergometry with Swan-Ganz catheterization and spiroergometry.
Results:
The automatic, closed-loop speed control showed robust and stable performance. It provided an increase in pump flow (+0.94 +/- 0.5 liters/min, p < 0.05) compared with constant-speed mode in response to physical activity. Pulmonary arterial (PAP) and capillary wedge pressure (PCWP) clearly decreased (-7.4 +/- 4.1 mm Hg for PAP and -8.3 +/- 4.2 mm Hg for PCWP, p < 0.05), and venous oxygen saturation moderately increased (+5.2%).
Conclusion:
An automatic speed-control system for rotary blood pumps was developed and demonstrated by spiroergometry to be appropriately responsive to physiologic demand.
More Related Videos
08:51Multiple Intravenous Bolus Dosing and Invasive Hemodynamic Assessment in a Hypoxia-Induced Mouse Pulmonary Artery Hypertension Model
Published on: November 11, 2022
09:13Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization I: Pre-Procedure Overview
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Cardiac Catheterization IV: Nursing Management
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...