An interchangeable Mapleson A-E breathing system is practical and cost effective
1Department of Anesthesia, Stanford University School of Medicine and Palo Alto Veterans Affairs Health Care System, Palo Alto, CA 94304, USA.
Modifying a Mapleson D breathing system to Mapleson A reduces fresh gas flow (FGF) requirements during anesthesia. This modification offers potential cost and resource savings in low-resource settings.
Area of Science:
- Anesthesiology
- Respiratory System Engineering
- Medical Device Modification
Background:
- Limited oxygen and anesthesia gas supplies necessitate reduced fresh gas flow (FGF) during inhalational anesthesia.
- Circle systems are often impractical in resource-limited environments.
- Investigating modifications to transport breathing apparatus for improved gas efficiency is crucial.
Purpose of the Study:
- To evaluate the interchangeability between Mapleson D (Map-D) and Mapleson A (Map-A) configurations using a common transport breathing apparatus.
- To determine if the Map-A configuration can eliminate carbon dioxide (CO2) rebreathing at lower FGF rates compared to Map-D.
Main Methods:
- A standard Map-D transport system was converted to a Map-A system by repositioning the exhaust valve and fresh gas delivery point.
- Comparison of Map-A and Map-D systems was performed under a structured protocol.
- Key metrics included FGF, FGF/minute ventilation (V(E)), and physiological variables.
Main Results:
- The Map-A configuration achieved a mean decrease in FGF of 2.8 L/min compared to Map-D (P=0.003).
- FGF/V(E) was significantly lower with Map-A (1.1) versus Map-D (1.8) (P=0.007) without affecting physiologic or anesthetic variables.
- The reduction in FGF achievable by switching systems correlated with patient respiratory rate (r=0.45, P<0.01).
Conclusions:
- Modification of breathing systems to achieve Mapleson A configuration can significantly reduce fresh gas flow requirements.
- This modification offers potential for cost and resource savings in clinical settings.
- Appropriate ventilation can be maintained at lower FGF rates, enhancing efficiency.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)