Related Experiment Video
Updated: Aug 11, 2026

Direct and Indirect Culture Methods for Studying Biodegradable Implant Materials In Vitro
Published on: April 15, 2022
[The study on biocompatibility of diamond-like carbon coated nickel-titanium shape memory alloy with osteoblasts
Qiang Li1, Qin Zhang, Zhengyi Sun
1Institute of Orthopaedics, Second Hospital of Lanzhou University, Lanzhou Gansu 730030, PR China.
Objective:
To investigate the biocompatibility of diamond-like carbon (DLC) coated Nickel-Titanium shape memory alloy with osteoblasts cultured in vitro.
Methods:
Rabbit's osteoblasts were incubated with DLC-coated Nickel-Titanium shape memory alloy disks and uncoated ones of equal size for 5 days. The control group (without shape memory alloy in culture media) was performed simultaneously. The cultured cells were counted and graphed. The samples from culture media were collected and the concentrations of alkaline phosphatase (ALP) and nickel (Ni2+) were measured from the 1st to 5th day respectively.
Results:
The proliferation of osteoblasts and the concentration of ALP in both DLC-coated group and control group was higher than uncoated group. The proliferation of osteoblasts on the 3rd, 4th, and 5th day in both DLC-coated group and control group was significantly higher than that in the uncoated group (P < 0.05). The concentration of ALP in DLC-coated group on the 2nd, 3rd, and 5th day and in the control group on the 3rd, 4th, and 5th day was significantly higher than that in the uncoated group (P < 0.05). The concentration of Ni2+ on the 3rd, 4th, and 5th day was significantly lower than that in the uncoated group (P < 0.05).
Conclusion:
DLC- coated Nickel-Titanium shape memory alloys appears to have better biocompatibility with osteoblast cultured in vitro compared to uncoated ones.