High throughput quantification of cholesterol and cholesteryl ester by electrospray ionization tandem mass
Gerhard Liebisch1, Marion Binder, Rainer Schifferer
1Institute for Clinical Chemistry and Laboratory Medicine, University of Regensburg, Germany.
Abstract:
Analysis of free cholesterol (FC) is not well suited for electrospray ionization (ESI); however, cholesteryl ester (CE) form ammonium adducts in positive ion mode and generate a fragment ion of m/z 369 upon collision-induced fragmentation. In order to allow parallel analysis of FC and CE using ESI tandem mass spectrometry (ESI-MS/MS), we developed an acetyl chloride derivatization method to convert FC to cholesteryl acetate (CE 2:0). Derivatization conditions were chosen to provide a quantitative conversion of FC to CE 2:0 without transesterification of naturally occurring CE species. FC and CE were analyzed by direct flow injection analysis using a fragment of m/z 369 in a combination of selected reaction monitoring (SRM) and precursor ion scan for FC and CE, respectively. Quantification was achieved using deuterated D(7)-FC and CE 17:0/CE 22:0 as internal standards as well as calibration lines generated by addition of FC and naturally occurring CE species to the respective sample matrix. The developed assay showed a precision and detection limit sufficient for routine analysis. A run time of 1.3 min and automated data analysis allow high throughput analysis. Loading of human skin fibroblast and monocyte derived macrophages with stable isotope labeled FC showed a potential application of this method in metabolism studies. Together with existing mass spectrometry methodologies for lipid analysis, the present methodology will provide a useful tool for clinical and biochemical studies and expands the lipid spectrum that can be analyzed from one lipid sample on a single instrumental platform.
More Related Videos
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Tandem Mass Spectrometry
Chemical Ionization (CI) Mass Spectrometry