X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Subjective and objective measures of image quality in digital fluoroscopy
1Department of Medical Physics and Bio-Engineering, St James's Hospital, Dublin 8, Ireland. clwalsh@stjames.ie
Abstract:
There are numerous methods for assessing image quality in diagnostic X ray. In our study we assessed how imaging quality assurance methods perform in practice. Physics assessments were based on IPEM protocols using Leeds test objects. Clinical assessment was based on a questionnaire. A total of 15 systems in three European locations were assessed, covering a range of image intensifier-TV digital fluoroscopy units. Analysis of 274 clinical questionnaires showed that clinical and physics assessments did not place systems in the same order, based on a given image quality parameter. In almost all the comparisons, low level correlation was measured for statistical comparison of rank order (rs < 0.3). However, broad agreement was observed between physics and clinical assessments for image quality associated with contrast and noise. This study emphasises the importance of maintaining links with clinical assessment, when developing quality assurance metrics, and measuring the mutual performance of clinical and physical assessments of image quality.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
X-ray Imaging
Imaging Studies III: Computed Tomography