Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Induction therapy of pralsetinib followed by conversion surgery for unresectable locally advanced non-small cell lung cancer with the rearranged during transfection (RET) fusion gene mutation: protocol for a single-arm phase II prospective trial (the SALVAGE-P trial).

Journal of thoracic disease·2026
Same author

How I Do It-Intra-Operative Needle Puncture Marking for Wedge Resection of Deep Posterobasal Lung Nodules.

ANZ journal of surgery·2026
Same author

Paeoniflorin as a candidate disease-modifying therapy for diabetic peripheral neuropathy: mechanisms, exposure challenges, and translational priorities.

Frontiers in pharmacology·2026
Same author

L-shaped Reduction Malarplasty: Its Geometric Consideration.

Aesthetic plastic surgery·2026
Same author

Variation and fitness of hybrid F<sub>4</sub> between glyphosate-tolerant soybeans and three wild soybean populations from Northern China.

Pest management science·2026
Same author

A Method for Locating the Summit of Zygoma and Radiographic Measurement of Zygoma in Patients With Hemifacial Microsomia.

The Journal of craniofacial surgery·2026

Related Experiment Video

Updated: Apr 11, 2026

Multi-Scale Modification of Metallic Implants With Pore Gradients, Polyelectrolytes and Their Indirect Monitoring In vivo
12:19

Multi-Scale Modification of Metallic Implants With Pore Gradients, Polyelectrolytes and Their Indirect Monitoring In vivo

Published on: July 1, 2013

11.3K

[Scanning electron microscopic study of titanium-hydroxyapatite, the functionally graded-material implanted in

Xiao-jun Tang1, Shu-zhong Xing, Xiao-ling Song

  • 1Plastic Surgery Hospital, Beijing, China.

Zhonghua Zheng Xing Wai Ke Za Zhi = Zhonghua Zhengxing Waike Zazhi = Chinese Journal of Plastic Surgery
|February 9, 2006
PubMed
Summary

This study compared how well different implant materials integrate with bone tissue in rabbits. The materials tested were a titanium-hydroxyapatite functionally graded material, pure titanium, and pure hydroxyapatite. Researchers used scanning electron microscopy to observe new bone formation around each material at 2, 4, and 8 weeks after implantation. They found that the titanium-hydroxyapatite graded material formed a direct bond with new bone by two months, while pure titanium left a small gap. The graded material showed better early integration than pure titanium and similar results to pure hydroxyapatite. These findings suggest that titanium-hydroxyapatite functionally graded materials may be more effective for implants than pure titanium.

Keywords:
functionally graded materialsosseointegrationtitanium implantshydroxyapatiteveterinary biomaterials

Frequently Asked Questions

More Related Videos

Biological Compatibility Profile on Biomaterials for Bone Regeneration
10:28

Biological Compatibility Profile on Biomaterials for Bone Regeneration

Published on: November 16, 2018

13.5K
Calvarial Model of Bone Augmentation in Rabbit for Assessment of Bone Growth and Neovascularization in Bone Substitution Materials
08:41

Calvarial Model of Bone Augmentation in Rabbit for Assessment of Bone Growth and Neovascularization in Bone Substitution Materials

Published on: August 13, 2019

9.1K

Related Experiment Videos

Last Updated: Apr 11, 2026

Multi-Scale Modification of Metallic Implants With Pore Gradients, Polyelectrolytes and Their Indirect Monitoring In vivo
12:19

Multi-Scale Modification of Metallic Implants With Pore Gradients, Polyelectrolytes and Their Indirect Monitoring In vivo

Published on: July 1, 2013

11.3K
Biological Compatibility Profile on Biomaterials for Bone Regeneration
10:28

Biological Compatibility Profile on Biomaterials for Bone Regeneration

Published on: November 16, 2018

13.5K
Calvarial Model of Bone Augmentation in Rabbit for Assessment of Bone Growth and Neovascularization in Bone Substitution Materials
08:41

Calvarial Model of Bone Augmentation in Rabbit for Assessment of Bone Growth and Neovascularization in Bone Substitution Materials

Published on: August 13, 2019

9.1K

Area of Science:

  • Biomaterials in orthopedic surgery
  • Tissue integration research in veterinary medicine

Background:

Tissue integration of implants remains a key challenge in biomaterials science. Prior research has shown that pure titanium and hydroxyapatite each have distinct advantages in osseointegration. However, the behavior of functionally graded materials combining these components is less understood. No prior work had resolved how these hybrid materials interact with bone tissue over time. This gap motivated the current investigation into Ti-HA functionally graded materials. The study aimed to clarify how these materials compare with pure titanium and pure hydroxyapatite in terms of tissue response. Researchers needed to determine if the graded structure influences integration differently than single-material implants. Understanding these interactions could improve implant design in veterinary and human applications. The early-stage tissue response is particularly important for predicting long-term success.

Purpose Of The Study:

The study aimed to evaluate how Ti-HA functionally graded materials interact with bone tissue compared to pure titanium and pure hydroxyapatite. Researchers wanted to determine if the graded structure promotes better integration than single-component materials. The specific problem addressed was the lack of data on how Ti-HA FGM behaves in a living system. The motivation came from the need to improve implant integration in clinical settings. By comparing three materials in a controlled model, the study sought to identify optimal integration characteristics. The rabbit parietal bone was selected as a suitable model for observing osseointegration. The focus was on early-stage tissue response and interface formation. The goal was to inform future biomaterial design and implant development.

Main Methods:

The study involved implanting three types of materials into rabbit parietal bones. The materials tested were sintered Ti-HA FGM, pure hydroxyapatite, and pure titanium. Each material was implanted in separate groups of rabbits. Specimens were retrieved at 2, 4, and 8 weeks post-surgery for analysis. Scanning electron microscopy was used to examine tissue-material interfaces. Researchers assessed new bone formation around each implant type. The focus was on comparing the amount and maturity of new bone across groups. Interface characteristics were analyzed to determine integration quality.

Main Results:

At two weeks, Ti-HA FGM showed more new bone formation than pure titanium. The maturity of new bone surrounding Ti-HA FGM was also higher than pure titanium. At four weeks, the pattern remained similar to the two-week results. By eight weeks, direct bonding between Ti-HA FGM and new bone was observed. Pure titanium implants showed a small gap between the material and new bone. Pure hydroxyapatite exhibited integration comparable to Ti-HA FGM. The tissue tolerance of Ti-HA FGM was confirmed through consistent findings. These results suggest Ti-HA FGM integrates with bone more effectively than pure titanium.

Conclusions:

The authors reported that Ti-HA FGM integrates with bone more effectively than pure titanium. The material-bone interface formed as an integral structure two months post-implantation. The tissue tolerance of Ti-HA FGM was confirmed through SEM observations. Pure hydroxyapatite showed similar integration to Ti-HA FGM. The findings suggest Ti-HA FGM may be a suitable implant material. The early-stage integration of Ti-HA FGM was better than pure titanium. The results support the potential use of functionally graded materials in implants. These conclusions are based on the observed interface characteristics and tissue response.

The study found that Ti-HA FGM forms a direct bone interface two months post-implantation, unlike pure titanium which leaves a small gap.

Scanning electron microscopy was used to observe bone formation around implants at 2, 4, and 8 weeks post-surgery.

The parietal bone was selected as a suitable model for observing osseointegration in a living system.

SEM was used to examine the structure and maturity of new bone surrounding each implant type.

Ti-HA FGM showed integration similar to pure hydroxyapatite in terms of new bone formation and interface quality.

The authors suggest Ti-HA FGM may be a suitable material for implants due to its superior integration compared to pure titanium.