Clinical comparison of tympanic membrane displacement with invasive ICP measurements
S Shimbles1, C Dodd, K Banister
1Regional Medical Physics Department, Newcastle General Hospital, Newcastle upon Tyne, UK.
Objective:
Tympanic membrane displacement (TMD) measurements may be useful in the management of patients with hydrocephalus if they can be directly associated with measurements of ICP. We have compared TMD measurements using the Marchbanks Measurement System with invasive ICP monitoring.
Methods:
Twenty-nine patients who were undergoing routine invasive monitoring using a Camino fibre optic ICP measurement system as part of their clinical management were studied. Simultaneous measurements of ICP and TMD in both sitting and supine positions were successfully made in thirteen patients.
Results:
Thirty-nine pairs of readings were obtained. The invasive ICP readings varied from 1 to 36 mmHg in the supine position and from -12 to +35 mmHg sitting. Corresponding TMD values varied from 275 to +277 nL in the supine position and -133 to +466 nL sitting. Linear regression showed a significant negative relationship between the two measurements (r = -0.57, p = 0.0013).
Conclusions:
There is a strong negative linear association between mean TMD and invasively measured ICP and this relationship is highly significant. Nevertheless, TMD is a poor surrogate for ICP in clinical terms because the predictive limits of the linear regression are too wide. However, serial intra-patient measurements may be useful to determine changes in ICP with time.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Increased Intracranial Pressure l: Introduction