Related Experiment Video
Updated: Aug 11, 2026

Short-Duration Hypothermia Induction in Rats using Models for Studies examining Clinical Relevance and Mechanisms
Published on: March 3, 2021
Experimental study on a large animal model of a new thermoablation technique
C Roux1, N Rauber, E Hiltbrand
1Centre d'Etudes et de Recherches Médicales d'Archamps, Domaine de Chosal, 74160 Archamps, France.
Background:
A novel technique of thermoablation, using a microtube to deliver pulses of hot water vapour, was tested on a large animal model in order to evaluate its efficacy and potential adverse effects.
Materials And Methods:
The medical device consisted of a microtube extension connected to a hydropneumatic pump. Pulses of pure water were injected though the microtube where they were heated and delivered as vapour into the target zone. The method was tested on the liver of 12 healthy pigs, either during open surgery or percutaneously under ultrasounds.
Results:
The technique was efficient and well-tolerated by the animals. Large volumes of necrotic tissue were created in a significantly short time compared to concurrent thermoablative techniques.
Conclusion:
Anticipating human application, this experimental study demonstrated a safe and efficient innovative thermoablation technique. The first human applications have been successfully performed and will be reported soon.