An improved analytical model for CT dose simulation with a new look at the theory of CT dose
Robert L Dixon1, Michael T Munley, Ersin Bayram
1Department of Radiology, Wake Forest University School of Medicine, Medical Center Boulevard, Winston-Salem, North Carolina 27157-1088, USA. rdixon@wfubmc.edu
Abstract:
Gagne [Med. Phys. 16, 29-37 (1989)] has previously described a model for predicting the sensitivity and dose profiles in the slice-width (z) direction for CT scanners. The model, developed prior to the advent of multidetector CT scanners, is still widely used; however, it does not account for the effect of anode tilt on the penumbra or include the heel effect, both of which are increasingly important for the wider beams (up to 40 mm) of contemporary, multidetector scanners. Additionally, it applied only on (or near) the axis of rotation, and did not incorporate the photon energy spectrum. The improved model described herein transcends all of the aforementioned limitations of the Gagne model, including extension to the peripheral phantom axes. Comparison of simulated and measured dose data provides experimental validation of the model, including verification of the superior match to the penumbra provided by the tilted-anode model, as well as the observable effects on the cumulative dose distribution. The initial motivation for the model was to simulate the quasiperiodic dose distribution on the peripheral, phantom axes resulting from a helical scan series in order to facilitate the implementation of an improved method of CT dose measurement utilizing a short ion chamber, as proposed by Dixon [Med. Phys. 30, 1272-1280 (2003)]. A more detailed set of guidelines for implementing such measurements is also presented in this paper. In addition, some fundamental principles governing CT dose which have not previously been clearly enunciated follow from the model, and a fundamental (energy-based) quantity dubbed "CTDI-aperture" is introduced.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Pharmacodynamic Models: Linear Concentration–Effect Model
Pharmacokinetic Models: Overview
There are three primary types of models: empirical, compartment, and physiological. Empirical models, with minimal assumptions,...