Analytical performance and clinical efficacy for cardiovascular risk estimation of an Olympus immunoturbidimetric
Snezana Jovicić1, Svetlana Ignjatović, Marijana Dajak
1Institute of Medical Biochemistry, Clinical Center of Serbia, Belgrade, Serbia and Montenegro.
Insights
The Olympus hsCRP assay demonstrates good analytical performance and clinical efficacy for cardiovascular disease risk assessment, showing strong agreement with a validated comparative method. This finding supports its use in estimating cardiovascular risk in apparently healthy adults.
Area of Science:
- Clinical Chemistry
- Cardiovascular Medicine
- Biomarker Analysis
Background:
- Elevated C-reactive protein (CRP) within the normal range (<10.0 mg/L) predicts cardiovascular disease (CVD) risk in healthy adults.
- High-sensitivity CRP (hsCRP) assays are crucial for accurate CVD risk stratification.
- Cutoff values (<1, 1-3, >3 mg/L) define low, average, and high CVD risk.
Purpose of the Study:
- To evaluate the analytical performance of the Olympus immunoturbidimetric latex CRP assay (sensitive application).
- To assess the clinical efficacy of the Olympus hsCRP assay for cardiovascular risk estimation.
- To compare the Olympus assay against the FDA-approved CardioPhase* hsCRP assay.
Main Methods:
- Assessed imprecision of the Olympus hsCRP assay (0.38-8.16% within-run, 3.75-9.63% between-run) in the 0.2-10.0 mg/L range.
- Conducted method comparison using 194 serum samples (0.15-10.0 mg/L CRP) with weighted Deming regression and Bland-Altman analysis.
- Evaluated agreement in relative risk class assignments between Olympus and Dade Behring methods.
Main Results:
- Olympus hsCRP assay showed excellent analytical performance with high precision.
- Weighted Deming regression yielded a slope of 0.99 mg/L and r=0.992, indicating strong correlation.
- Agreement in CVD risk classification was 95.4%, with no statistically significant differences between methods (p>0.10).
Conclusions:
- The Olympus hsCRP assay exhibits good analytical performance and clinical efficacy for CVD risk assessment.
- The assay demonstrates reliable agreement with the established CardioPhase* hsCRP method.
- This study validates the Olympus hsCRP assay for estimating cardiovascular risk, contrary to some previous findings.
Abstract:
Increased C-reactive protein (CRP) concentration within the reference interval (<10.0 mg/L) is a strong predictor of cardiovascular disease (CVD) in apparently healthy adults. Cutoff points for use of CRP in estimating CVD risk are <1, 1-3 and >3 mg/L for low, average and high relative risk, respectively. For measuring CRP concentrations to assess cardiovascular risk, high-sensitivity CRP (hsCRP) assays have been developed. The aim of this study was to evaluate the analytical performance and clinical efficacy for cardiovascular risk estimation of the Olympus immunoturbidimetric latex CRP assay (sensitive application). The comparative method used was the CardioPhase* hsCRP assay, approved by the Food and Drug Administration for use in CVD risk assessment. The imprecision of the Olympus hsCRP assay in the concentration range 0.2-10.0 mg/L was 0.38-8.16% within runs and 3.75-9.63% between runs. For method comparison studies, 194 fresh serum samples were selected to cover the interval 0.15-10.0 mg/L CRP. Comparison of the Dade Behring and Olympus methods was performed using weighted Deming regression analysis (slope 0.99 mg/L, intercept 0.002 mg/L, S(y,x)=0.02 mg/L, r=0.992) and a Bland-Altman relative difference plot (mean difference -0.002%, SD=0.040%). The agreement between the Dade Behring and Olympus methods for relative risk class assignments was 95.4%. Statistical analysis of the agreement between the two methods for each relative risk class showed that the differences between the methods were not statistically significant (p>0.10). Although previous reports found poor performance of the Olympus CRP tests for use in cardiovascular and peripheral vascular risk estimation, our study proved good analytical performance and clinical efficacy of the Olympus hsCRP assay for this use.
Related Concept Videos
Blood Studies for Cardiovascular System II: CRP, Hcy, and Cardiac Natriuretic Peptide Markers
These markers indicate stress or strain on the heart muscle:
Natriuretic Peptides (BNP)
Cardiac myocytes produce these hormones in response to ventricular stretching...
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT