Dosimetric evaluation of a new OneDose MOSFET for Ir-192 energy
Rajesh A Kinhikar1, Pramod K Sharma, Chandrashekhar M Tambe
1Department of Medical Physics, Tata Memorial Hospital, Parel, Mumbai 400 012, India. rkinhikar@rediffmail.com
Abstract:
The purpose of this study was to investigate dosimetry (reproducibility, energy correction, relative response with distance from source, linearity with threshold dose, rate of fading, temperature and angular dependence) of a newly designed OneDosetrade mark MOSFET patient dosimetry system for use in HDR brachytherapy with Ir-192 energy. All measurements were performed with a MicroSelectron HDR unit and OneDose MOSFET detectors. All dosimeters were normalized to 3 min post-irradiation to minimize fading effects. All dosimeters gave reproducible readings with mean deviation of 1.8% (SD 0.4) and 2.4% (SD 0.6) for 0 degrees and 180 degrees incidences, respectively. The mean energy correction factor was found to be 1.1 (range 1.06-1.12). Overall, there was 60% and 40% mean response of the MOSFET at 2 and 3 cm, respectively, from the source. MOSFET results showed good agreement with TLD and parallel plate ion chamber. Linear dose response with threshold voltage shift was observed with applied doses of 0.3 Gy-5 Gy with Ir-192 energy. Linearity (R2 = 1) was observed in the MOSFET signal with the applied dose range of 0.3 Gy-5 Gy with Ir-192 energy. Fading effects were less than 1% after 10 min and the MOSFET detectors stayed stable (within 5%) over a period of 1 month. The MOSFET response was found to be decreased by approximately 1.5% at 37 degrees C compared to 20 degrees C. The isotropic response of the MOSFET was found to be within +/-6%. A maximum deviation of 5.5% was obtained between 0 degrees and 180 degrees for both the axes and this should be considered in clinical applications. The small size, cable-less, instant readout, permanent storage of dose and ease of use make the MOSFET a novel dosimeter and beneficial to patients for skin dose measurements with HDRBT using an Ir-192 source compared to the labour demanding and time-consuming TLDs.
Related Concept Videos
Biological Effects of Radiation
MOSFET: Enhancement Mode
In their basic form, enhancement-mode MOSFETs are typically non-conductive when the gate-source voltage (Vgs) is zero. This default 'off' state means no current...
Characteristics of MOSFET
Various vital parameters influence their functionality, which is crucial for theory and electronics applications. First, channel dimensions, precisely length, and width, are pivotal. The size of these channels affects the transistor's ability to carry current and switching speeds; shorter channels typically enable quicker...
Isotopes and Radioisotopes
An isotope containing more...
MOSFET: Depletion Mode
The primary characteristic of depletion-mode MOSFETs is their ability to conduct current between the drain and source terminals without gate bias. This inherent conductivity arises...