Related Experiment Video
Updated: Aug 11, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Monte Carlo characterization of an ytterbium-169 high dose rate brachytherapy source with analysis of statistical
David C Medich1, Mark A Tries, John J Munro
1Radiation Laboratory, University of Massachusetts Lowell, 1 University Avenue, Lowell, Massachusetts 01854, USA. david_medich@uml.edu
Abstract:
An ytterbium-169 high dose rate brachytherapy source, distinguished by an intensity-weighted average photon energy of 92.7 keV and a 32.015 +/- 0.009 day half-life, is characterized in terms of the updated AAPM Task Group Report No. 43 specifications using the MCNP5 Monte Carlo computer code. In accordance with these specifications, the investigation included Monte Carlo simulations both in water and air with the in-air photon spectrum filtered to remove low-energy photons below 10 keV. TG-43 dosimetric data including S(K), D(r, lamda), lambda, gL(r), F(r, lamda), phi an(r), and phi(an) were calculated and statistical uncertainties in these parameters were derived and calculated in the appendix.