Related Experiment Video
Updated: Aug 11, 2026

Assessing Early Stage Open-Angle Glaucoma in Patients by Isolated-Check Visual Evoked Potential
Published on: May 25, 2020
Influence of test reliability on the screening performance of frequency-doubling perimetry
Govert P Heeg1, Nomdo M Jansonius
1Department of Ophthalmology, University Medical Center Groningen, University of Groningen, 9700 RB Groningen, The Netherlands.
Purpose:
To investigate the influence of test reliability on the screening performance of frequency-doubling perimetry (FDT).
Design:
Cross-sectional study.
Methods:
FDT sensitivity and specificity were calculated three times using three different strategies for handling unreliable (that is, >0 reliability indices outside normal limits) and unfeasible (that is, could not be completed) tests in 452 glaucoma patients from our outpatient department and 237 healthy subjects recruited outside the hospital.
Results:
Unreliable (P = .01) and unfeasible (P < .001) tests occurred more frequently in glaucoma patients when compared with healthy subjects. Best screening performance was obtained by interpreting unreliable tests as if they were reliable and considering unfeasible tests as positive. With this strategy, sensitivity was 0.90 (95% confidence interval 0.87-0.93) and specificity 0.81 (95% confidence interval 0.76-0.86).
Conclusions:
Unreliable and unfeasible FDT tests appear to contain useful information.