Related Experiment Video
Updated: Aug 11, 2026

A Chemical Screening Procedure for Glucocorticoid Signaling with a Zebrafish Larva Luciferase Reporter System
Published on: September 10, 2013
A comparison of assay performance measures in screening assays: signal window, Z' factor, and assay variability ratio
Philip W Iversen1, Brian J Eastwood, G Sitta Sittampalam
1Lilly Research Laboratories, Eli Lilly & Company, Indianapolis, Indiana; 303 W. Broadway St. Decorah, IA 52101, USA. piversen@lilly.com
Abstract:
In this article, the authors compare the assay performance measures, signal window, Z' factor, and assay variability ratio. They examine their mathematical formulae for similarities and differences, describe their statistical sampling properties using the results of a computer simulation, and illustrate their use with example data. Based on these results, the authors recommend the Z' factor as a preferred measure of assay performance for screening assays and point out that none of these measures are adequate for characterizing concentration-response assays.