Variability in a commercially available enzyme-linked immunosorbent assay system. II. Laboratory variability
D L Kreider1, J K Skeeles, M Parsley
1Department of Animal and Poultry Sciences, University of Arkansas, Fayetteville 72701.
Avian Diseases
|April 1, 1991
Summary
Variability in enzyme-linked immunosorbent assays (ELISA) is significant across different laboratories and days. Single serum sample tests are unreliable; standard reference pools are needed for accurate antibody titer estimation.
Area of Science:
- Veterinary Immunology
- Diagnostic Assays
- Laboratory Science
Background:
- Enzyme-linked immunosorbent assays (ELISA) are widely used for antibody titer determination.
- Inter-laboratory and intra-laboratory variability can impact the reliability of ELISA results.
- Standardization is crucial for accurate and reproducible diagnostic testing in poultry health.
Purpose of the Study:
- To quantify the variability in ELISA results across multiple laboratories and days.
- To assess the reliability of single serum sample determinations for antibody titer.
- To determine the necessity of standardized reference pools in poultry diagnostic assays.
Main Methods:
- Serum pools of varying antibody titers (low, medium, high) were distributed to five cooperating poultry industry laboratories.
- Assays were conducted over three consecutive days at each laboratory.
- Data on antibody titers were collected and analyzed for inter-laboratory and intra-laboratory variations.
Main Results:
- Significant variations in ELISA results were observed among different laboratories.
- Substantial variability was also found among different testing days within the same laboratory.
- The greatest source of variability was identified as differences among laboratories.
- Correlations between daily mean titer and laboratory temperature were not significant.
Conclusions:
- Single determinations on individual serum samples are not reliable for estimating antibody titers due to high variability.
- The significant variability within and among laboratories highlights the critical need for standard reference pools.
- Implementing standard reference pools is essential to substantiate ELISA assay results and improve diagnostic accuracy.
More Related Videos
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
In 1971, Peter Perlman and Eva Engvall developed an Enzyme-linked immunosorbent assay (ELISA or EIA). ELISA differs from western blot in that the assays are conducted in microtiter plates or in vivo rather than on an absorbent membrane.
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Automated Microbial Diagnostics
Automated diagnostic analyzers have transformed clinical microbiology by providing rapid and reliable methods for pathogen identification and antibiotic susceptibility testing. Among these systems, the Vitek 2 is widely used because it automates the traditionally labor-intensive processes of microbial identification (ID) and antibiotic susceptibility testing (AST), delivering standardized and timely results that are essential for effective patient care.Microbial Identification with ID CardsThe...