Related Experiment Video
Updated: Aug 11, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac
Koichi Chida1, Haruo Saito, Hiroki Otani
1Department of Radiological Technology, Tohoku University School of Health Sciences, 2-1 Seiryo-machi, Aoba-ku, Sendai 980-8575, Japan. chida@rad.cms.tohoku.ac.jp
Objective:
Real-time maximum dose monitoring of the skin is unavailable on many of the X-ray machines that are used for cardiac intervention procedures. Therefore, some reports have recommended that physicians record the fluoroscopic time for patients undergoing fluoroscopically guided intervention procedures. However, the relationship between the fluoroscopic time and the maximum radiation skin dose is not clear. This article describes the correlation between the maximum radiation skin dose and fluoroscopic time for patients undergoing cardiac intervention procedures. In addition, we examined whether the correlations between maximum radiation skin dose and body weight, fluoroscopic time, and dose-area product (DAP) were useful for estimating the maximum skin dose during cardiac intervention procedures.
Materials And Methods:
Two hundred consecutive cardiac intervention procedures were studied: 172 percutaneous coronary interventions and 28 cardiac radiofrequency catheter ablation (RFCA) procedures. The patient skin dose and DAP were measured using Caregraph with skin-dose-mapping software.
Results:
For the RFCA procedures, we found a good correlation between the maximum radiation skin dose and fluoroscopic time (r = 0.801, p < 0.0001), whereas we found a poor correlation between the maximum radiation skin dose and fluoroscopic time for the percutaneous coronary intervention procedures (r = 0.628, p < 0.0001). There was a strong correlation between the maximum radiation skin dose and DAP in RFCA procedures (r = 0.942, p < 0.0001). There was also a significant correlation between the maximum radiation skin dose and DAP (r = 0.724, p < 0.0001) and weight-fluoroscopic time product (WFP) (r = 0.709, p < 0.0001) in percutaneous coronary intervention procedures.
Conclusion:
The correlation between the maximum radiation skin dose with DAP is more striking than that with fluoroscopic time in both RFCA and percutaneous coronary intervention procedures. We recommend that physicians record the DAP when it can be monitored and that physicians record the fluoroscopic time when DAP cannot be monitored for estimating the maximum patient skin dose in RFCA procedures. For estimating the maximum patient skin dose in percutaneous coronary intervention procedures, we also recommend that physicians record DAP when it can be monitored and that physicians record WFP when DAP cannot be monitored.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Pharmacokinetic–Pharmacodynamic Relationship: Duration of Dose-Effect Relationship
Biological Effects of Radiation
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Drug Accumulation During Multiple Dosing: Repetitive IV Injections