64-slice computed tomography assessment of coronary artery stents: a phantom study
A H Mahnken1, G Mühlenbruch, T Seyfarth
1Department of Diagnostic Radiology, Aachen University of Technology, Germany. mahnken@rad.rwth-aachen.de
Purpose:
To compare the use of a new 64-slice computed tomography (CT) scanner with 16-slice CT in the visualization of coronary artery stent lumen.
Material And Methods:
Eight different coronary artery stents, each with a diameter of 3 mm, were placed in a static chest phantom. The phantom was positioned in the CT gantry at an angle of 0 degrees and 45 degrees towards the z-axis and examined with both a 64-slice and a 16-slice CT scanner. Effective slice thickness was 0.6 mm with 64-slice CT and 1 mm with 16-slice CT. A reconstruction increment of 0.3 mm was applied in both scanners. Image quality was assessed visually using a 5-point grading scale. Stent diameters were measured and compared using paired Wilcoxon tests.
Results:
Artificial lumen reduction was significantly less with 64-slice than with 16-slice CT. Average visible stent lumen was 53.4% using 64-slice CT and 47.5% with 16-slice MSCT. Most severe artifacts were seen in stents with radiopaque markers. Using 64-slice CT, image noise increased by approximately 30% due to thinner slice thickness.
Conclusion:
Improved spatial resolution of 64-slice CT resulted in superior assessment of coronary artery stent lumen compared to 16-slice CT. However, a relevant part of the stent lumen is still not assessable with multi-slice CT.
More Related Videos
06:39Novel Percutaneous Approach for Deployment of 3D Printed Coronary Stenosis Implants in Swine Models of Ischemic Heart Disease
Published on: February 18, 2020
09:57Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT