Related Experiment Video
Updated: Aug 11, 2026

Surgical Model for Single-Staged Tissue-Engineered Urothelial Tubes in Minipigs
Published on: July 5, 2024
Novel everting urologic access sheath: potential advantages of decreased cellular advancement
Affonso H L A Camargo1, Jonathan N Rubenstein, Sinan Sozen
1Department of Urology, University of California San Francisco, San Francisco, California 94143, USA.
Background And Purpose:
Axial forces are imposed on the urothelium during advancement of instruments across the urinary tract, potentially transferring cellular debris, bacteria, or urothelial carcinoma from one anatomic location to another. A prototype access sheath (Cystoglide; Percutaneous Systems, Mountain View, CA) was created that everts and radially dilates but does not provide axial forces during deployment that can be used in a variety of anatomic systems. We created a urinary-tract model to evaluate the in-vitro advancement of cells to compare this technology with using instruments alone.
Materials And Methods:
Blocks of sterile agar were created with 17F tracts of three lengths (2.7, 5.5, and 11 cm) with 5 mL of Luria-Bertani broth/ampicillin solution in a well at the end. The tips of a Cystoglide sheath and a traditional urologic instrument of the same diameter were dipped into a suspension of ampicillin-resistant Escherichia coli and advanced through the tracts. After a 10-second exposure, 4 mL of broth was collected and cultured. Bacterial growth was compared by measuring the optical density (OD) of the broth at multiple time points.
Results:
The mean overall OD of the broth was significantly lower (P < 0.001) in the novel-sheath cultures than with a traditional instrument for all advancements at all tract lengths.
Conclusion:
The Cystoglide sheath significantly reduces the advancement of cells within an artificial urinary tract compared with a non-everting instrument. Clinical studies are needed to assess the utility of this technology in vivo.