Related Experiment Video
Updated: Aug 11, 2026

Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Performance evaluation of the immature granulocyte parameter on the Sysmex XE-2100 automated hematology analyzer
Diane Field1, Elizabeth Taube, Sherry Heumann
1St. Anthony's Medical Center, St. Louis, Missouri 63128, USA. fieldj@samcstl.org
Abstract:
Recently, St. Anthony's Medical Center Laboratory (SAMC) in St. Louis, Missouri, installed a new automated hematology analyzer, the Sysmex XE-2100. The XE-2100 has an additional differential parameter, the immature granulocyte (IG) parameter. Prior to implementation, the IG parameter was evaluated by comparison to the manual immature granulocyte count. Immature granulocytes counted manually, as well as the automated count, included metamyelocytes, myelocytes, and promyelocytes. Manual differentials were performed on each specimen by 2 experienced medical technologists using the National Committee for Clinical Laboratory Standards (NCCLS) document H20-A. In addition to evaluating the IG parameter, St. Anthony's laboratory plans to report the automated IG count from the XE-2100, with
