Related Experiment Video
Updated: Aug 11, 2026

Gas Chromatography-Mass Spectrometry Paired with Total Vaporization Solid-Phase Microextraction as a Forensic Tool
Published on: May 25, 2021
Quantification of clotiazepam in human plasma by gas chromatography-mass spectrometry
Sung-Hoon Ahn1, Han-Joo Maeng, Tae-Sung Koo
1Department of Pharmaceutics, College of Pharmacy, Seoul National University, San 56-1, Shinlim-dong, Kwanak-gu, 151-742, South Korea.
Abstract:
An analytical procedure was developed and validated for the quantification of clotiazepam in human plasma. After subjecting plasma samples to solid-phase extraction, the extract was evaporated and the residue re-constituted. An aliquot of the mixture was injected onto a gas chromatography-mass spectrometry system. The detector response was linear for clotiazepam concentrations in the range of 5-200 ng/ml. Intra- and inter-day precision for the assay over the concentration range was below 13.1 and 13.5%, and the accuracy ranged between 99.0-107.9% and 92.4-101.3%, respectively. The drug was found to be stable under various processing conditions used. The method is applicable to human pharmacokinetic studies of clotiazepam.
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...