Related Experiment Video
Updated: Aug 10, 2026

A Novel Platform for In Vitro Cellular Stretching and Imaging
Published on: March 10, 2026
A novel device for in vitro isometric tension recordings of cylindrical artery segments
Joseph B Young1, Babak S Jahromi, Zhen-Du Zhang
1Section of Neurosurgery, MC3026, Department of Surgery, University of Chicago Medical Center and Pritzker School of Medicine, 5841 South Maryland Avenue, Chicago, IL 60637, USA.
Abstract:
There are few instruments specifically designed to measure circumferential force generated by ring segments of arteries in vitro. Typical limitations of some existing machines include poor isometry, large organ bath volume or difficult sample mounting. The authors designed, built and tested a device for isometric tension recording of force developed by rings of arteries in vitro. It is suitable for assessment of arteries from 0.3 to 3 mm in diameter and allows measurements of forces in the range 0-20 g on eight rings simultaneously. The organ baths are independently regulated in temperature, stirred, disposable and have a minimum useable volume of only 1.2 mL.