Related Experiment Video
Updated: Aug 10, 2026

An Automated Method for Assessing Visual Acuity in Infants and Toddlers Using an Eye-Tracking System
Published on: March 17, 2023
Accuracy and reliability of remote retinopathy of prematurity diagnosis
Michael F Chiang1, Jeremy D Keenan, Justin Starren
1Department of Ophthalmology, Columbia University College of Physicians and Surgeons, New York, NY, USA. chiang@dbmi.columbia.edu
Objective:
To determine the accuracy and reliability of retinopathy of prematurity (ROP) diagnosis using remote review of digital images by 3 masked ophthalmologist readers.
Methods:
An atlas was compiled of 410 retinal photographs from 163 eyes of 64 low-birth-weight infants taken using a wide-angle digital fundus camera. All the images were independently reviewed by 3 readers, and the diagnosis in each eye was classified into 1 of 4 ordinal categories: no ROP, mild ROP, type 2 prethreshold ROP, or ROP requiring treatment. Findings were compared with a reference standard of dilated indirect ophthalmoscopy with scleral depression performed by an experienced pediatric ophthalmologist.
Results:
Sensitivities/specificities of the diagnosis of any ROP were 0.845/0.910 for the first reader, 0.816/0.955 for the second reader, and 0.864/0.493 for the third reader. Sensitivities/specificities of the diagnosis of ROP requiring treatment were 0.850/0.960 for the first reader, 0.850/0.973 for the second reader, and 0.900/0.953 for the third reader. When ROP was classified into ordinal categories, the overall weighted kappa for interreader reliability was 0.743. Intrareader reliability for detection of low-risk prethreshold ROP or worse was 100% for all readers.
Conclusion:
The accuracy, interreader reliability, and intrareader reliability of remote diagnosis of clinically relevant ROP based on digital imaging are substantial.