Related Experiment Video
Updated: Aug 10, 2026

Direct Comparison of Hyperspectral Stimulated Raman Scattering and Coherent Anti-Stokes Raman Scattering Microscopy for Chemical Imaging
Published on: April 28, 2022
Chemical imaging by single pulse interferometric coherent anti-stokes Raman scattering microscopy
Sang-Hyun Lim1, Allison G Caster, Olivier Nicolet
1Department of Chemistry, University of California at Berkeley, California 94720, USA.
Abstract:
A single pulse interferometric coherent anti-Stokes Raman (CARS) spectroscopy method is used to obtain broadband CARS spectra and microscopy images of liquid and polymer samples. The pump, Stokes, and probe pulses are all selected inside a single broadband ultrafast pulse by a phase- and polarization-controlled pulse shaping technique and used to generate two spectral interference CARS signals simultaneously. The normalized difference of these two signals provides an amplified background-free broadband resonant CARS spectrum over the 400-1500 cm(-1) range with 35 cm(-1) spectral resolution. Chemically selective microscopy images of multicomponent polymer and liquid samples are investigated with this new CARS method. Multiplex CARS spectra at 10,000 spatial points are measured within a few minutes, and used to construct chemically selective microscopy images with a spatial resolution of 400 nm. The spectral bandwidth limits, sensitivity, homodyne amplification advantages, spatial resolution, depolarization, chromatic aberration, and chemical imaging aspects of this new technique are discussed in detail.