Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
Published on: September 19, 2017
Sensitive electrochemical enzyme immunoassay microdevice based on architecture of dual ring electrodes with a sensing
1School of Chemical and Biomedical Engineering, Nanyang Technological University, Nanyang Avenue, Singapore 639798, Singapore.
Abstract:
A novel electrochemical detection architecture was investigated for enzyme immunoassay sensors. Microchips with dual-ring working and counter electrodes, and a sensing cavity chamber were made on glass slides. The glass surface of the microchip was coated by 3-aminopropyltriethoxysilane (APTES). Goat IgG, as a example, was covalently captured on APTES-modified glass surfaces through glutaraldehyde (GA) as a cross-linker. Enzyme substrate, p-aminophenyl phosphate (PAPP) was prepared by electrolysis. The enzyme conversion from home-synthetic PAPP to p-aminophenol (PAP) was examined by differential pulse voltammetry (DPV). A competitive inhibition enzyme-linked immunosorbant assay (ELISA) was designed to test the system. Experimental results demonstrate that a detection limit of 118 fg/ml of goat IgG and a dynamic range of 118 fg/ml to 1.18 ng/ml, up to five orders of magnitude could be achieved. Due to its novel architecture design and electronic detection scheme, the method can be used to fabricate portable electrochemical ELISA lab-on-chip systems. The technology could have great potential in clinical diagnostic applications.
Related Concept Videos
Potentiometry: Membrane Electrodes
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.