Sheathless Capillary Electrophoresis–Mass Spectrometry for Metabolic Profiling of Biological Samples
Published on: October 1, 2016
Copper-coated microsprayer interface for on-line sheathless capillary electrophoresis electrospray mass spectrometry
Alina D Zamfir1, Nicolae Dinca, Eugen Sisu
1Institute for Medical Physics and Biophysics, University of Münster, Germany. zamfir@uni-muenster.de
Abstract:
A sturdy home-built sheathless CE/ESI-QTOF-MS system was developed and optimized for carbohydrate analysis. The interface and employed methodology provided a simple analytical solution to laborious CE/MS interfacing methods and to problems in characterization of complex carbohydrate mixtures that require high-resolution separation of the components. The CE/ESI interface, feasible in any MS laboratory, consists of a one-piece CE column having the CE terminus in-laboratory shaped as a microsprayer and coated with copper. The CE microsprayer was inserted into an in-house made stainless steel clenching device and the whole assembly was mounted onto a quadrupole TOF mass spectrometer. The analytical potential of the interface in terms of suitability, microsprayer performance, copper coat durability, ionization efficiency, spray stability, and sensitivity was tested first on a simple mixture of standard saccharides, which were separated, resolved, and detected with high separation efficiency. The approach was next assessed for the screening of a biological sample, a complex mixture of O-glycosylated sialylated amino acids from urine of a patient suffering from Schindler disease. Preliminary data allow this method to be considered as one of general applicability in structural glycobiology and glycomics and easy to be implemented for proteomic surveys as well.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
07:54Capillary Electrophoresis Mass Spectrometry Approaches for Characterization of the Protein and Metabolite Corona Acquired by Nanomaterials
Published on: October 27, 2020
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Capillary Electrophoresis: Instrumentation
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview