Optimization by factorial design of a capillary zone electrophoresis method for the simultaneous separation of
M Elisa Capella-Peiró1, Alessandra Bossi, Josep Esteve-Romero
1Dipartimento Scientifico e Tecnologico, Università degli Studi di Verona, Strada Le Grazie 15, 37134 Verona, Italy. capella@guest-mail.uji.es
Abstract:
A 3(2) full factorial design was used to optimize the experimental conditions of a capillary zone electrophoresis method aimed at achieving simultaneous separation and quantification of the antihistamines brompheniramine, chlorpheniramine, cyproheptadine, diphenhydramine, doxylamine, hydroxyzine, and loratadine according to their therapeutic group. A statistical program, SPSS, was used to calculate the mathematical model with which to obtain the response surface. Critical parameters such as pH and applied voltage were studied to evaluate their effect on resolution and on efficiency. Optimum separation conditions were phosphate buffer pH 2.0, 5kV, and 2psis(-1) at 214nm. The analysis time was below 9min and the theoretical plates were between 6000 and 63,000N. Calibration curves were prepared for the antihistamines. The limits of detection were 4-14ngmL(-1), which allow their quantification in pharmaceuticals. The RSD% of each antihistamine was fairly good. Up to seven antihistamines belonging to the antihistaminic H(1)-receptor group were separated in the same electropherogram. The proposed method was then applied to the determination of antihistamines in pharmaceutical, urine, and serum samples with recoveries in agreement with the stated contents.
More Related Videos
Related Concept Videos
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Capillary Electrophoresis: Instrumentation
High-Performance Liquid Chromatography: Elution Process