Suppression of ionization and optimization of assay for 3-hydroxy fatty acids in house dust using ion-trap mass
K Udeni Alwis1, Lennart Larsson, Donald K Milton
1Department of Environmental Health, Exposure Epidemiology and Risk Program, Harvard School of Public Health, Boston, Massachusetts, USA.
Background:
3-Hydroxy fatty acids (3-OHFAs), components of lipid A of gram-negative bacteria are useful chemical markers of endotoxin.
Methods:
We analyzed 3-OHFAs in house dust as trimethylsilyl (TMS) methyl ester derivatives in the electron impact ionization mode using gas chromatography ion-trap mass spectrometry. Linear calibrations with r > 0.995 were observed for all the 3-OHFA methyl ester external standards.
Results:
Recovery efficiency experiments with house dust demonstrated that accurate quantification requires calibration curves to be subjected to phase separation and solid phase extraction (SPE) because of differing clean-up losses according to chain length of 3-OHFAs. Recovery experiments also demonstrated interference with detection of C16:0 by the ion trap, which may be due to suppression of ionization by a constituent in house dust. Interference was overcome by injecting 1:4 dilutions of derivatized samples. The range of recoveries was 89.3%-111.5% for 3-OHFAs added to house dust. The reproducibility of injections was high (CV for C14:0 = 2.36%). The limit of detection (LOD) was 0.15 ng/mg for each 3-OHFA.
Conclusions:
The modifications we made included: use of 3-hydroxy C11:0 and C13:0 methyl esters as internal standards, subjecting calibration standards to phase separation and SPE; addition of water to phase separation; addition of 1-pentadecanol as a carrier; injecting 1:4 diluted TMS derivatives of 3-OHFAs; and monitoring both m/z 131 and 133 ions to improve stability of area measurements for product ions. This method of optimization establishes an appropriate technique for quantification of 3-OHFAs in house dust.
More Related Videos
08:23Analyzing the Photo-oxidation of 2-propanol at Indoor Air Level Concentrations Using Field Asymmetric Ion Mobility Spectrometry
Published on: June 14, 2018
05:52Histone Modification Screening using Liquid Chromatography, Trapped Ion Mobility Spectrometry, and Time-Of-Flight Mass Spectrometry
Published on: January 12, 2024
Related Concept Videos
Mass Analyzers: Common Types
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Chemical Ionization (CI) Mass Spectrometry