Related Experiment Video
Updated: Aug 9, 2026

Quantitative Immunofluorescence to Measure Global Localized Translation
Published on: August 22, 2017
Measurement and restoration of the point spread function of fluorescence confocal microscopy
1Nano Opto-Mechatronics Laboratory, Department of Mechanical Engineering, Korea Advanced Institute of Science and Technology, Daejeon 305-701, South Korea. turtle99@kaist.ac.kr
Abstract:
The point spread function of an objective lens of a fluorescence confocal microscope was directly measured by imaging fluorescent beads. We analysed how the measurement of the point spread function was influenced by the diameter of the fluorescent beads and how the restoration technique with a deconvolution algorithm improved the measuring performance. Numerical and experimental results are presented for a typical point spread function and a zero-centred point spread function.
More Related Videos
Related Concept Videos
Super-resolution Fluorescence Microscopy
Confocal Fluorescence Microscopy