Related Experiment Video
Updated: Jul 11, 2026

Construction and Implantation of a Microinfusion System for Sustained Delivery of Neuroactive Agents.
Published on: March 18, 2008
Infusion pump delivers over-dosage of propofol as a result of missing syringe support
Christian Koch1, Christine Hollister, Peter H Breen
1Department of Anesthesiology, University of California-Irvine, Orange, California, USA.
Abstract:
We describe the malfunction of a common drug infusion pump. The syringe saddle was missing and allowed the syringe barrel to contact the pump case, which decreased the outward displacement of the syringe clamp. Then, the infusion pump falsely detected a smaller syringe size and consequently delivered an increased infusion rate and overdose of propofol to the patient. More commonly, an incorrectly mounted syringe may increase the outward displacement of the syringe clamp so that the infusion pump falsely detects a larger syringe size, with resultant less than expected infusion rate.