Verification imaging using a computed radiography system for high-energy electron beam therapy.

Hideki Fujita1, Michihiro Yamaguchi, Toshizo Katsuda

  • 1Department of Radiology, Osaka Prefectural Medical Center for Respiratory and Allergic Diseases, Habikino 583-8588, Japan.

Radiation Medicine
|March 25, 2006
PubMed
Summary

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Treatment Guidance for Palmoplantar Pustulosis 2022.

The Journal of dermatology·2026
Same author

Eye lens dose distribution in neonatal head CT: an analysis of the scanning conditions and CTDIvol limitations.

Radiation protection dosimetry·2026
Same author

Directional Dependence of Gafchromic LD-V1.

Journal of medical physics·2026
Same author

Comparison of Severity Assessment Methods in Generalized Pustular Psoriasis: A Single-Center Study.

The Journal of dermatology·2026
Same author

5-hydroxyeicosatetraenoic acid and 12-hydroxyeicosapentaenoic acid regulate basophil and mast cell activation and basophil recruitment in chronic spontaneous urticaria.

Allergology international : official journal of the Japanese Society of Allergology·2026
Same author

Response to comments on "Effectiveness of transitioning from omalizumab to dupilumab in chronic spontaneous urticaria patients with inadequate response to omalizumab".

The World Allergy Organization journal·2025

Computed radiography (CR) systems with diagnostic cassettes provide convenient, high-quality verification images for radiation therapy quality assurance. This method is clinically useful and suitable for daily use.

Area of Science:

  • Medical Imaging
  • Radiation Oncology
  • Radiography

Background:

  • Verification imaging is crucial for quality assurance in radiation therapy.
  • Current methods for acquiring verification images can be inconvenient.
  • Computed radiography (CR) offers a potential alternative for image acquisition.

Purpose of the Study:

  • To evaluate a convenient method for acquiring verification images using a CR system.
  • To assess the image quality and clinical utility of verification images obtained with a CR system and diagnostic cassette.

Main Methods:

  • Verification images were acquired using both diagnostic and therapy cassettes with a CR system.
  • Image processing parameters were optimized to enhance image quality.
  • Radiation technicians and a radiation oncologist evaluated image quality and clinical usefulness.

Related Experiment Videos

Main Results:

  • Optimized image processing significantly improved verification image quality.
  • No significant difference in image quality was found between diagnostic and therapy cassettes (p>0.05).
  • Verification images were deemed useful in 98% of clinical cases.

Conclusions:

  • A CR system utilizing a diagnostic cassette can produce high-quality verification images.
  • This CR system is practical for daily verification imaging in radiation therapy.
  • The system is considered valuable for maintaining quality assurance in high-energy electron beam therapy.