Effect of coupling medium temperature on rate of intramuscular temperature rise using continuous ultrasound
C A Oshikoya1, S J Shultz, D Mistry
1Denver Broncos Football Club, Englewood, CO.
Objective:
We determined the effects of coupling medium temperature on the rate of intramuscular temperature rise (RTR) during continuous ultrasound.
Design And Setting:
Ultrasound was applied in a continuous mode at a frequency of 1 MHz and intensity of 1.5 W/cm(2). Each subject received 3 treatments, using water-based coupling gel at temperatures of 18 degrees C, 25 degrees C, and 39 degrees C. All treatments were performed in an athletic training room under controlled environmental conditions.
Subjects:
Eighteen healthy male subjects (mean age = 23.6 +/- 3.5 years; height = 177.8 +/- 6.9 cm; weight = 76.6 +/- 8.2 kg; calf size = 37.6 +/- 2.4 cm) participated in this study.
Measurements:
A thermistor was inserted into the left medial triceps surae at a depth of 5 cm, and baseline tissue temperatures were recorded before treatment. Intramuscular temperature was recorded every 30 seconds until the temperature rose 4 degrees C above baseline or until discomfort was felt. RTR was calculated by dividing the absolute temperature change by treatment time.
Results:
A 1-way, repeated-measures analysis of variance revealed a significant difference in RTR among gel temperatures, RTR was significantly faster using the 25 degrees C gel compared with the 18 degrees C and 39 degrees C gels. There was no difference between the 18 degrees C and 39 degrees C gel treatments.
Conclusions:
These results suggest that the use of a cooled or heated gel may be counterproductive when maximal thermal effects are desired within a given time frame.
More Related Videos
08:523D Ultrasound Imaging: Fast and Cost-effective Morphometry of Musculoskeletal Tissue
Published on: November 27, 2017
13:41Magnetic Resonance-Guided High Intensity Focused Ultrasound Generated Hyperthermia: A Feasible Treatment Method in a Murine Rhabdomyosarcoma Model
Published on: January 13, 2023
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Muscle Stimulation Frequency
Wave summation
At low firing rates, motor neurons induce individual twitch contractions in muscle fibers. These twitches...