Finite element simulations of tip-enhanced Raman and fluorescence spectroscopy
Andrew Downes1, Donald Salter, Alistair Elfick
1Institute of Materials and Processes, University of Edinburgh, King's Buildings, Mayfield Road, Edinburgh EH9 3JL, United Kingdom. andy.downes@ed.ac.uk
Abstract:
Finite element electromagnetic simulations of scanning probe microscopy tips and substrates are presented. The enhancement of the scattered light intensity is found to be as high as 10(12) for a 20 nm radius gold tip, and tip-substrate separation of 1 nm. Molecular resolution imaging (< 1 nm) is achievable, even with a relatively large radius tip (20 nm). We also make predictions for imaging in aqueous environments, noting a sizable red shift of the spectral peaks. Finally, we discuss signal levels, and predict that high-speed Raman mapping should be possible with gold substrates and a small tip-substrate separation (< 4 nm).
Related Concept Videos
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...