Single breath-hold T1 measurement using low flip angle TrueFISP.

L Bokacheva1, A J Huang, Q Chen

  • 1Department of Radiology, New York University School of Medicine, New York, New York 10016, USA.

Magnetic Resonance in Medicine
|March 31, 2006
PubMed
Summary

This study introduces a fast T1 mapping method using segmented inversion recovery TrueFISP (sIR-TrueFISP) for single breath-hold scans. The technique accurately measures T1 values in phantoms and human tissues, proving valuable for abdominal and chest imaging.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

[Single non-blood-related umbilical cord blood transplantation using a reduced-intensity conditioning regimen for the treatment of severe aplastic anemia].

Zhonghua xue ye xue za zhi = Zhonghua xueyexue zazhi·2024
Same author

Choroid Plexus Calcification Correlates with Cortical Microglial Activation in Humans: A Multimodal PET, CT, MRI Study.

AJNR. American journal of neuroradiology·2023
Same author

[Spectrum of gene mutations and clinical features in adult acute myeloid leukemia with normal karyotype].

Zhonghua xue ye xue za zhi = Zhonghua xueyexue zazhi·2022
Same author

[Rituximab-induced acute thrombocytopenia in a patient with splenic marginal zone lymphoma: a case report and literature review].

Zhonghua xue ye xue za zhi = Zhonghua xueyexue zazhi·2021
Same author

Bronchiectasis is associated with delayed diagnosis and adverse outcomes in the New Zealand Common Variable Immunodeficiency Disorders cohort study.

Clinical and experimental immunology·2021
Same author

Preliminary Findings Associate Hippocampal <sup>1</sup>H-MR Spectroscopic Metabolite Concentrations with Psychotic and Manic Symptoms in Patients with Schizophrenia.

AJNR. American journal of neuroradiology·2020

Area of Science:

  • Magnetic Resonance Imaging (MRI)
  • Biomedical Engineering
  • Medical Physics

Background:

  • Accurate T1 quantification is crucial for various MRI applications, including tissue characterization and contrast agent assessment.
  • Traditional T1 mapping techniques often require long acquisition times, limiting their clinical utility, especially in abdominal and thoracic imaging due to respiratory motion.

Purpose of the Study:

  • To develop and validate a novel, rapid T1 estimation method using a single breath-hold, segmented inversion recovery prepared, true fast imaging with steady-state precession (sIR-TrueFISP) acquisition.
  • To compare the accuracy and reliability of the proposed sIR-TrueFISP technique against a standard T1 mapping method.

Main Methods:

  • Implementation of a low flip angle (FA) sIR-TrueFISP sequence for T1 mapping.

Related Experiment Videos

  • Acquisition of T1 measurements in a Gd-DTPA-doped water phantom and in the brain and abdomen of healthy volunteers.
  • Comparison of sIR-TrueFISP derived T1 values with those obtained using the conventional inversion recovery fast spin echo (IR-FSE) technique.
  • Main Results:

    • Excellent correlation (R²=0.999) between sIR-TrueFISP and IR-FSE in the phantom.
    • Good correlation (R²=0.943) between the two methods in human brain and abdominal tissues.
    • T1 values obtained using sIR-TrueFISP were in good agreement with previously published data.

    Conclusions:

    • The sIR-TrueFISP technique provides accurate T1 measurements within a single breath-hold.
    • This rapid T1 mapping method is highly suitable for clinical applications, particularly for imaging organs affected by physiological motion such as the chest and abdomen.