Related Experiment Video
Updated: Aug 9, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Pharmacokinetic study of orphenadrine using high-performance liquid chromatography-tandem mass spectrometry
Soo-Youn Lee1, Hyeon Ju Oh, Jong Won Kim
1Department of Laboratory Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea. suddenbz@smc.samsung.co.kr
Abstract:
We developed and validated a simple, rapid, and accurate HPLC-MS/MS method with simple protein precipitation for the determination of orphenadrine. Injection-to-injection running time was 3 min with a retention time of orphenadrine of 1.1 min. The linear assay range was 1-200 ng/mL (r2 > 0.99). The intra- and inter-assay imprecisions were CV 0.6-4.2% and CV 1.6-6.1%, respectively. The accuracy, extraction recovery, specificity and stability were satisfactory. Using the measured plasma concentrations of orphenadrine in 24 healthy subjects, pharmacokinetic profiles of orphenadrine were evaluated (AUC(0-72,) 1565+/-731 ng h/mL, Cmax 82.8+/-26.2 ng/mL, Tmax 3.0+/-0.9 h, elimination half-life 25.8+/-10.3 h).
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Tandem Mass Spectrometry
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...