Varying MLC end projection size: an effect of the double-focused Siemens multileaf collimator
1Regional Medical Physics Department, Newcastle General Hospital, Westgate Road, Newcastle upon Tyne NE4 6BE, UK. andrew.buckle@nuth.nhs.uk
Abstract:
Siemens linacs use multileaf collimators (MLCs) that move in an arc such that the flat faces of the leaf ends are always in the same plane as the radiation focus. An effect of this is that the magnification of the leaf end projection at the isocentric plane normal to the collimator rotation axis varies with the drive position of the leaves. This effect is quantified theoretically and empirically verified. A method is introduced for assessing the importance of the effect for a particular MLC pattern. The significance of the effect is discussed.