High-frequency mechanical vibration to recanalize chronic total occlusions after failure to cross with conventional

Eberhard Grube1, Gabor Sütsch, Victor Y Lim

  • 1Klinikum Siegburg, Siegburg, Germany.

Insights

The CROSSER system safely crosses chronic total occlusions (CTOs) when standard guidewires fail. This innovative device shows promise for improving interventional cardiology success rates in complex coronary artery disease cases.

Area of Science:

  • Interventional Cardiology
  • Medical Device Technology
  • Cardiovascular Research

Background:

  • Chronic total occlusions (CTOs) present significant challenges in coronary artery interventions.
  • Conventional guidewire techniques often fail to cross these complex lesions.
  • The CROSSER system, using high-frequency mechanical vibration, was developed to address this unmet need.

Purpose of the Study:

  • To evaluate the feasibility and safety of the CROSSER system for crossing CTOs.
  • To determine the effectiveness of the CROSSER system in patients who failed conventional guidewire crossing.
  • To assess the primary endpoint of device efficacy and secondary endpoint of clinical success.

Main Methods:

  • A two-phase clinical study involving 53 patients and 55 CTO lesions across four European centers.
  • Phase 1 focused on device safety in 30 CTOs.
  • Phase 2, the pivotal study, assessed effectiveness in 25 CTOs using an improved CROSSER device.

Main Results:

  • The pivotal study phase demonstrated a 76% success rate for both device efficacy and clinical success (19 of 25 CTOs).
  • No major adverse cardiac events, including death, myocardial infarction, or target lesion revascularization, were reported within 30-day follow-up.
  • The study reported no instances of coronary perforation or pericardial tamponade, highlighting the device's safety profile.

Conclusions:

  • The CROSSER system is a safe and effective device for crossing challenging coronary artery chronic total occlusions after conventional guidewire failure.
  • The system shows potential to improve procedural success rates in complex coronary interventions.
  • Further device refinement and improvements may lead to even higher success rates in future applications.
Abstract

Related Concept Videos