Related Experiment Video

Updated: Aug 9, 2026

The Intra-Aortic Balloon Pump
06:13

The Intra-Aortic Balloon Pump

Published on: February 5, 2021

Initial experiences of removal of intra-aortic balloon pumps with the Angio-Seal

Koji Kato1, Naoki Sato, Takeshi Yamamoto

  • 1Intensive and Cardiac Care Unit, Nippon Medical School, Tokyo, Japan.

Insights

Angio-Seal effectively and safely achieved hemostasis after intra-aortic balloon pump removal in 16 patients. No major or minor complications were observed within seven days, demonstrating its utility in vascular hemostasis.

Area of Science:

  • Vascular Surgery
  • Interventional Cardiology

Background:

  • Angio-Seal is a proven vascular hemostasis device for cardiac catheterization.
  • Its efficacy in intra-aortic balloon pump (IABP) removal is not well-established.

Purpose of the Study:

  • To evaluate the safety and effectiveness of Angio-Seal for hemostasis after IABP removal.

Main Methods:

  • A prospective study involving 16 patients undergoing IABP removal.
  • Angio-Seal device was used for achieving hemostasis.
  • Safety endpoint included major (e.g., retroperitoneal bleeding, vessel occlusion) and minor (e.g., hematoma, pseudoaneurysm) complications within 7 days.

Main Results:

  • Rapid hemostasis was achieved, with a mean time of 3.3 minutes (range 2-5 minutes).
  • No major complications were recorded.
  • No minor complications, including hematoma, AV fistula, or pseudoaneurysm, were observed within the 7-day follow-up period.

Conclusions:

  • Angio-Seal is a safe and effective device for achieving rapid hemostasis following IABP removal.