Is the body-coil at 3 Tesla feasible for the MRI evaluation of the painful knee? A comparative study
G Lutterbey1, K Behrends, M V Falkenhausen
1Department of Radiology, University of Bonn, Sigmund-Freud-Strasse 25, 53105 Bonn, Germany. goetz.lutterbey@ukb.uni-bonn.de
Abstract:
The purpose of this study was to compare the in-built body coil of the 3.0-Tesla (T) scanner with a dedicated surface coil of a 1.5 T system regarding knee imaging. We performed an intraindividual prospective clinical trial on 17 patients with knee pain using magnetic resonance imaging (MRI) at 1.5 and 3.0 T systems equipped with identical gradient systems. Proton-density-weighted turbo spin echo sequences with the same spatial resolution and comparable contrast parameters were used. A quantitative measurement of signal to noise ratio (SNR), relative contrast (RC) and contrast to noise ratio (CNR) between muscle and bone marrow was performed, followed by a qualitative assessment of anatomic/pathologic structures and the extent of artefacts. At 3.0 T, 30 lesions (91%) compared to 33 lesions at 1.5 T were detected. The SNR/CNR/RC were moderately reduced at 3.0 T versus 1.5 T (muscle 42 vs 47 and bone 83 vs 112/46 vs 69/0.33 vs 0.43). Motion artefacts from the pulsating popliteal artery were significantly increased at 3.0 T. A visible and measurable signal loss occurred at 3.0 T using the built-in body coil compared with the dedicated 1.5 T knee coil, but nearly all clinically important information could be obtained.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...