Dosimetry for Cell Irradiation using Orthovoltage (40-300 kV) X-Ray Facilities
Published on: February 20, 2021
Energy and angular anisotropy optimisation of a p-type diode for in vivo dosimetry in photon-beam radiotherapy
1Physics Department, Clatterbridge Centre for Oncology NHS Trust, Bebington, Wirral, UK.
Abstract:
We present simulation work using the Monte Carlo code MCNPX that shows that there is a possibility of improving the silicon p-type diode as a radiation dosemeter, by altering the construction of the diode. Altering the diode die thickness can reduce the inherent angular anisotropy of the diode, with little effect on its energy response. Conversely, the contact material and geometry have a large impact on the energy response with little effect on the inherent angular anisotropy. By correct choice of contact material, the typical over-response -100 keV relative to the response at 60Co energy can be reduced from approximately 20 to 4. It is expected that further enhancements may be made with different geometries and materials.
More Related Videos
Related Concept Videos
Biasing of P-N Junction
In equilibrium, no external voltage is applied across the p-n junction. The depletion region is formed at the junction interface due to the diffusion of carriers, which leaves behind charged dopants, acceptors on the p-side, and donors on the n-side. These immobile charges create an electric field that prevents further diffusion of carriers. The related energy band...