Determination of natural and depleted uranium in urine at the ppt level: an interlaboratory analytical exercise
E A Ough1, B J Lewis, W S Andrews
1Department of Chemistry and Chemical Engineering, Royal Military College of Canada, 11 General Crerar Crescent, Kingston, ON, Canada. edough@telus.net
Abstract:
An analytical exercise was initiated in order to determine those procedures with the capability to measure total uranium and uranium (238U/235U) isotopic ratios in urine samples containing >0.02 microg U kg-1 urine. A host laboratory prepared six identical sets of twelve synthetic urine samples containing total uranium in the range of 25 to 770 ng U kg-1 urine and with 238U/235U isotopic ratios ranging from 138 (100% NU) to 215 (51% DU). Sets of samples were shipped to five testing laboratories (four based in Canada and one based in Europe). Each laboratory utilized one of the following analytical techniques: sector field inductively coupled plasma mass spectrometry (ICP-SF-MS), quadrupole inductively coupled plasma mass spectrometry (ICP-Q-MS), thermal ionization mass spectrometry (TIMS), and instrumental/delayed neutron activation analysis (I/DNAA), in their analyses.