Related Experiment Video
Updated: Aug 9, 2026

FRET Imaging in Three-dimensional Hydrogels
Published on: August 1, 2016
Optical imaging of dose distributions in Fricke gels
V Viti1, F d'Errico, M Pacilio
1Yale University School of Medicine, New Haven, CT 06510, USA. viti@iss.it
Abstract:
Ferrous-sulphate infused gels, or 'Fricke gels', encounter great interest in the field of radiation dosimetry, due to their potential for 3D radiation dose mapping. Typically, magnetic resonance (MR) relaxation rates are determined in these systems in order to derive the absorbed dose. However, when large concentration gradients are present, diffusion effects before and during the MR imaging may not be negligible. In these cases, optical techniques may represent a viable alternative. This paper describes research aimed at measuring 3D dose distributions in a Fricke-xylenol orange gel by measuring optical density with a CCD camera. This method is inexpensive and fast. A series of early experiments is described, in which optical density profiles were measured with a commercial microdensitometer for film dosimetry. The light box of the device was modified to work at 567 nm, close to the maximum absorbance of the ferric ion-xylenol orange complex. Under these conditions, the gel shows linearity with dose and high sensitivity.